Medical device manufacturers operate under FDA 21 CFR Part 820 (Quality System Regulation) and ISO 13485. Both require a documented CAPA (Corrective and Preventive Action) process. The 8D methodology maps directly to CAPA requirements and is accepted by FDA auditors as a structured problem-solving framework.
ISO 13485 Clause 8.5.2 Corrective Action and 8.5.3 Preventive Action require:
All of these align with 8D D0-D8.
1. **Risk Management Integration (ISO 14971)**: D7 must update the device Risk Management File, not just the FMEA. Every corrective action must be evaluated for new risk introduction
2. **Health Hazard Evaluation (HHE)**: Before D3 containment, conduct an HHE to determine if a field safety corrective action (recall) is required
3. **Complaint Handling (820.198)**: D2 problem description must reference the complaint record number and include patient/operator impact assessment
4. **Design History File (820.30)**: D6 corrective actions that change device design must update the DHF
5. **CAPA Timeline**: FDA expects 30-day acknowledgment, 60-day root cause, 90-day corrective action completion
**Problem**: Infusion pump displays Error Code E-07 (flow rate deviation) on 3 units from Lot #M2026-04.
1. Why E-07? → Flow sensor reading deviates >5% from setpoint after 200h of continuous operation
2. Why deviation? → Peristaltic pump tubing compression set exceeds 15%, reducing volumetric accuracy
3. Why compression set? → Tubing material (silicone) cured at 150°C instead of specified 165°C during supplier process
4. Why wrong temperature? → Supplier changed oven without notifying manufacturer — this is a supplier process change without customer notification
5. Why undetected? → Incoming inspection tested tubing dimensions but not cross-link density or compression set properties
**TRC**: Tubing compression set due to incorrect curing temperature
**MRC**: Supplier process change management failure; incoming inspection did not test functional properties at risk
**Risk Impact**: Flow rate deviation could cause over-infusion or under-infusion — potential patient harm